SMC

SMC FC2W520-04-3-X110 Automatic Water Flow Controller

Manufacturer part numberFC2W520-04-3-X110

Official SMC FC2W-X110 catalog (15-E663) opened 18 November 2026 lists FC2W520-04-3-X110 as a water flow controller: Karman-vortex detection, 2.0 to 16.0 L/min control range, 1.6 to 16.8 L/min measurement range, Rc 1/2 port, 3 m lead with connector, 0.2 to 0.4 MPa operating pressure at 0 to 50°C fluid temperature, 0.6 MPa withstand pressure, ±5% F.S. control accuracy, response 10 s or less, analogue voltage input 1 to 5 VDC, flow-rate pulse output, 24 VDC ±10%, and 530 g. Leakage when fully closed is 1 L/min or less; a shutoff function is not provided, and the catalog requires an external shutoff valve to stop flow in an emergency. Wetted materials listed: PPS, PP, POM, FKM, stainless steel 303, stainless steel 304, SCS13, non-grease. Applicable fluid is water; glycol compatibility is not inferred. Send the exact official part number and control-interface details to enquire. This page is not a stock list.

Rated flow control range
2.0–16.0 L/min 2
Process port
Rc 1/2 3
Working pressure
0.2–0.4 MPa at 0–50 °C fluid temperature; proof pressure 0.6 MPa 4
Control accuracy
±5% F.S. flow-control accuracy; response time ≤10 s 5
Request a Quote

SMC FC2W520-04-3-X110 Automatic Water Flow Controller

SMC FC2W520-04-3-X110 Automatic Water Flow Controller

Specifications

ParameterValue
SMC FC2W520-04-3-X110
Ordering codeFC2W520-04-3-X110; 1/2 in. Rc port; 3 m lead 1
Rated flow control range2.0–16.0 L/min 2
Process portRc 1/2 3
Working pressure0.2–0.4 MPa at 0–50 °C fluid temperature; proof pressure 0.6 MPa 4
Control accuracy±5% F.S. flow-control accuracy; response time ≤10 s 5
Power and signal24 VDC ±10%; analogue voltage input 1 to 5 VDC (official catalog); flow-pulse output 6
Medium and temperatureWater; fluid temperature 0–50 °C without freezing 7
Shutoff limitationController cannot fully close; an external shutoff valve is needed for emergency stop 8
Specifications
Official FC2W-X110 cataloghttps://ca01.smcworld.com/catalog/New-products-en/mpv/15-e663-fc2w/data/15-e663-fc2w.pdf (opened 18 November 2026; English catalog 15-E663; model column FC2W520-04-3-X110)
Official input signal (correction)24 VDC ±10%; analogue voltage input 1 to 5 VDC (official catalog); flow-pulse output 6
Official flow measurement range1.6 to 16.8 L/min 9
Official leakage when fully closed1 L/min or less 10
Official wetted materialsPPS, PP, POM, FKM, stainless steel 303, stainless steel 304, SCS13; non-grease specifications 11
Official weight530 g 12
Notes 12
  1. Exact catalogue ordering example
  2. medium: Water; catalogue specified operating conditions
  3. Confirm mating thread and line size in quote
  4. Proof pressure is not the normal working range
  5. Catalogue condition; not a fast-acting safety shutoff
  6. official catalog for FC2W520-04-3-X110; earlier 0–5 VDC wording was not the official input range
  7. Do not infer glycol or other coolant compatibility from water rating
  8. Functional boundary of FC2W-X110 series
  9. official catalog measurement range for FC2W520-04-3-X110; distinct from the 2.0 to 16.0 L/min control range
  10. official catalog note; a shutoff function is not provided. Use an external shutoff valve to stop flow in an emergency, as the catalog states.
  11. official catalog list for parts in contact with fluid
  12. official catalog value for FC2W520-04-3-X110