VOLKE · FJP
VOLKE FJP Integrated Flow, Pressure and Temperature Sensor
Manufacturer seriesFJP
An inline sensor combining flow, pressure and temperature measurement.
Specifications
| Parameter | Value |
|---|---|
| FJP | |
| Measured variables | Flow, pressure and temperature 1 |
| Flow measurement principle | Hall-sensed impeller 2 |
| Nominal diameter options | DN6, DN8, DN10, DN15, DN20, DN25 3 |
| Dn6 listed flow range | 0.5 to 10 L/min 4 |
| Dn15 listed flow range | 5 to 50 L/min 5 |
| Dn25 listed flow range | 12 to 120 L/min 6 |
| Thread options | G1/2 for DN6–DN15; G3/4 for DN20; G1 for DN25 7 |
| Temperature measurement range | 0 to 100 degC 8 |
| Supply voltage | 24 VDC 9 |
| Process interface material | 304 stainless steel 10 |
| Output choices | Three 4–20 mA channels or RS485 11 |
| Specifications | |
| Official product page | VOLKE FJP flow-temperature-pressure sensor, Shenzhen Volke Automation, opened 18 November 2026. Official product page: https://www.autovolke.com/index/marticle/index/aid/863.html |
| Official measured variables and principle | Flow, pressure and temperature in one body; official impeller with Hall-effect switch plus separate pressure and temperature elements |
| Official flow rows | DN6 0.5–10 L/min G1/2"; DN8 1–15 L/min G1/2"; DN10 3–30 L/min G1/2"; DN15 5–50 L/min G1/2"; DN20 8–80 L/min G3/4"; DN25 12–120 L/min G1". Official accuracy class 3–5 is manufacturer wording, not a YOCOOLING guarantee. |
| Official pressure / temperature | Pressure −0.1 to 1 MPa (official accuracy class 0.5); temperature 0 to 100 °C (official accuracy 1 °C) |
| Official electrical and materials | Outputs: three 4–20 mA channels or RS485; 24 VDC; official power consumption 3 W; housing aluminium oxide; process interface 304 SS; official IP67 (IP68 described as customizable) |
Notes 11
- single integrated housing; exact output version to be selected
- supplier product description; fluid properties and calibration affect measurement
- series size options, not one universal body or flow range
- DN6 row; stated flow accuracy grade 3–5 is not a guarantee for all fluids
- DN15 row; measurement accuracy conditions require controlled manual
- DN25 row; do not apply to smaller fittings
- manufacturer table; verify each order-code and mating seal
- measurement span, not a combined process-temperature or coolant compatibility rating
- supplier webpage summary; wiring variant must be confirmed
- interface only, not a complete wetted-material statement
- supplier webpage lists options; verify exact version, protocol and pinout before specifying